The symposium saw three speakers examine the current limitations on defining and delivering required dose distributions, with regard to their respective areas of expertise: external-beam X-ray radiotherapy, ion beam therapy and brachytherapy.
External-beam concerns
The discussion kicked off with a presentation from David Thwaites, director of the Institute of Medical Physics at the University of Sydney, Australia, who examined the situation for megavoltage X-ray radiotherapy. Thwaites pointed out that while sophisticated modern technologies are able to deliver almost anything, there remains a host of unanswered questions.
Firstly, do we actually know the appropriate dose distribution to deliver to an individual patient? Currently, said Thwaites, no, we don't, as we don't understand the radiobiology well enough on an individual patient basis. Multimodal and functional imaging can help, but limitations remain due to fusion uncertainties and less than optimal image quality. As such, much research effort is currently going into improving deformable image registration (DIR) techniques.
Inconsistent target delineation between observers can also cause difficulties. As can the fact that modern delivery techniques, such as intensity-modulated radiotherapy (IMRT), volumetric-modulated radiotherapy and image-guided radiotherapy, alter the dose-volume trade-offs, the constraints and the margins. For example, IMRT produces a different dose spread to previous delivery techniques, with larger low-dose regions, the implications of which are not yet fully understood.
The next question: "Can we model the required dose distribution accurately and identify the best treatment plan?" Just about, said Thwaites. "In terms of plan optimization, we need to balance quality versus delivery limitations and delivery efficiency," he explained. Other factors to consider include: whether to perform biological or physical optimization; whether uncertainty can be included in the planning process; and which delivery technique and plan is best for a particular case, using multi-criteria approaches. Accurate low-dose modelling remains a problem.
And once we know what dose to deliver, can we realistically and robustly deliver it? Yes we can, Thwaites said, with the caveat that it's vital to use image-guidance for complex treatments and to verify dose delivery. At present, most radiotherapy is performed using pre-treatment verification, but in vivo dosimetry could detect errors that may otherwise be missed. "We should be doing in vivo dosimetry with EPIDs," he explained. "This is not fully supported by manufacturers at the moment, but clearly, it's where we should be going."
Another issue that must be addressed is the ability to deal with inter- and intra-fractional variations. Adaptive radiotherapy, Thwaites said, is still very much in development, and can be dependent upon the DIR algorithm chosen. For larger motion changes, tracking systems are reported to work reasonably well, but are not yet at the point of widespread implementation. There is also a need for more efficient and integrated quality assurance tools and methods.
"Yes, we can deliver the dose distribution that we plan," Thwaites concluded. "But there are still significant uncertainties and significant research challenges. What's more, different developmental steps are inter-dependent – as one improves, it affects the others."
Proton observations
The next speaker to take the podium was Martijn Engelsman, from the Holland Particle Therapy Centre in Delft, the Netherlands. Engelsman addressed some of the issues surrounding proton and ion-beam therapy. While protons offer highly conformal dose delivery, he explained, the planned dose distribution is much more sensitive to set-up errors and anatomy variation.
Proton therapy is currently delivered using either pencil-beam scanning or passive scattering. Of the two, scanning offers higher target conformation, but is affected more by intra-fractional target motion. Intensity-modulated proton therapy (IMPT) offers the best conformity, but is most sensitive to anatomy variation.
"Delivering dose to a phantom works well, the machine is not the problem, you can deliver what you want," Engelsman explained. "But things become different in a patient." Clinical treatments introduce the need to deal with set-up errors, range uncertainties, and density changes that impact the beam range. For pencil-beam scanning, interplay effects between the patient's breathing motion and the beam motion also impact the delivery accuracy.
Engelsman described some of the countermeasures that are available for dealing with motion and interplay. For range uncertainties and set-up errors, options include robust optimization of treatment plans and, to a lesser extent, the use of safety margins. The move to robust optimization is technically solved, he noted, with work ongoing to analyse variation over time. However, it is not yet routinely understood or available in treatment planning systems.
Intrafraction motion during scanned proton therapy can be addressed using target repainting, an approach being actively investigated and pursued by multiple treatment centres but still a year or two away from routine use. Anatomy variations between fractions are best compensated for using adaptive therapy techniques. These include, for example, image guidance – although this is very new to proton therapy – and in vivo dosimetry such as PET and prompt gamma imaging.
Other options include breath-hold and gating-based treatments, which are already in clinical use, and tumour tracking, which is several years from routine application. The correct mix of range-uncertainty countermeasures, i.e. the dose distribution we want to deliver, is still an open question for most indications.
In conclusion, Engelsman pointed out that the best possible proton therapy is not yet routinely available for all cases. For example, while a solitary brain or spine lesion can be treated very well with proton therapy, applying IMPT to an extensive and moving lung tumour will prove challenging even for the most experienced and R&D supported proton therapy department. "We're going to get there, evolution pending," he told the ESTRO delegates. "But we're not there yet; proton therapy is not yet plug-and-play."
Thoughts on brachytherapy
The symposium's final speaker was Dimos Baltas from the Klinikum Offenbach in Germany, who discussed brachytherapy. He started by considering some analogies between external-beam radiotherapy and modern brachytherapy. For example: the beams are equivalent to the brachytherapy needle or catheter; beam shaping corresponds to the source stepping pattern; and raster scanning is equivalent using several catheters to cover a target volume.
The physics of dose delivery, however, differs between the two techniques. In brachytherapy, the source is so close to the target that the delivered dose is dominated by the source dwell time, and the dose distribution is more similar to the spot (Bragg peak) delivered by an ion beam. Brachytherapy is also affected by the interconnection between the source and the patient anatomy, unlike external-beam radiotherapy in which beam shaping occurs externally.
In basic terms, brachytherapy is performed by creating a pre-plan, implanting the applicator under image guidance, and then treating the patient. But while imaging patient anatomy prior to and during catheter implantation enables accurate source placement, beam's eye view imaging is not possible, and the catheter itself can alter the patient anatomy.
One problem arises because pre-treatment imaging is usually performed in separate room to the treatment suite. Baltas described a brachytherapy study comparing catheter positions at CT planning and during treatment. The researchers noted displacement in the few hours between the planning scan and treatment delivery and recommend verification of catheter positions immediately prior to treatment.
To do this, Klinikum Offenbach employs ultrasound imaging during brachytherapy delivery. Baltas noted that this results in a mean shift of anatomy and needles of as low as 1.0 mm. For high-dose-rate brachytherapy, this corresponds to dosimetric changes of less than 5%, implying that quality assurance measures must be implemented to achieve this 1 mm stability. It's also vital that there is no motion of the implant and anatomy prior to treatment delivery, for example, caused by removing the ultrasound probe.
Baltas rounded off his talk by considering the symposium's key question: "Can we deliver the dose distribution we plan?". Under specific conditions and scenarios, he said, data exist to support the answer "yes". However, he added, there's currently no technology available to routinely answer this for different localizations. There's also no way to measure the delivered fluence of a brachytherapy treatment; effort is needed to develop in vivo dosimetry that can perform such verification.
1 comment
Comments on this article are now closed.
Estimating the dose delivered elswhere...
Comments on this article are now closed.